Services on Demand
Journal
Article
Indicators
- Cited by SciELO
- Access statistics
Related links
- Cited by Google
- Similars in SciELO
- Similars in Google
Share
DYNA
Print version ISSN 0012-7353
Abstract
RUIZ-OLAYA, Andrés F.; CALLEJAS-CUERVO, Mauro and LARA-HERRERA, Claudia N.. Wearable low-cost inertial sensor-based electrogoniometer for measuring joint range of motion. Dyna rev.fac.nac.minas [online]. 2017, vol.84, n.201, pp.180-185. ISSN 0012-7353. https://doi.org/10.15446/dyna.v84n201.59054.
Human movement analysis have become fundamental for a wide range of applications, such as physical therapies, neuro-rehabilitation, sports medicine, human motor system performance assessment and occupational medicine. This paper focuses on the design, implementation and validation of an inertial sensor-based portable electrogoniometer aimed at measuring joint range of motion. The system is composed of two inertial measurement units and a microcontroller that permits real-time human motion capture. The electrogoniometer was tested both in static and dynamic conditions in comparison to a standard commercial electrogoniometer during elbow movements. For static validation, the concordance correlation coefficient (ρC) obtained was 0.9605. The ρC found for slow-speed cyclical movements was 0.9830 and the ρC found for moderate-speed cyclical movements was 0.9619.
Keywords : Electrogoniometer; inertial sensors; joint range of motion; motion capture system.