SciELO - Scientific Electronic Library Online

 
vol.38 issue1Analysis of tenso-deformacional behavior of unsaturated soil for circular foundation with axial loadReduction of opacity in pollutant gas emissions in compression ignition engines by using biodiesel B5 and B10 based on algae author indexsubject indexarticles search
Home Pagealphabetic serial listing  

Services on Demand

Journal

Article

Indicators

Related links

  • On index processCited by Google
  • Have no similar articlesSimilars in SciELO
  • On index processSimilars in Google

Share


Ingeniería y Desarrollo

Print version ISSN 0122-3461On-line version ISSN 2145-9371

Abstract

LEAN, PAUL; ALVAREZ, CARLOS; NORIEGA, AINSWORTH  and  NAVARRO, ERIK. Development and validation of the mechanical behavior of a cervical disc implant with an "S" shape of Ti-6AI-4V ELI manufactured by SLM. Ing. Desarro. [online]. 2020, vol.38, n.1, pp.197-211.  Epub June 11, 2021. ISSN 0122-3461.  https://doi.org/10.14482/inde.38.1.620.5.

In the present study the mechanical behavior of a titanium alloy cervical medical implant type Ti-6Al-4V ELI is developed and validated, manufactured by Selective Laser Melting (SLM), which is a highly versatile and a faster process that allows to manufacture components with complex shapes. The medical implant developed in this research presents improvements compared to existing ones. The proposal of this research is the geometric design optimization set up as an "S" shape, by using finite elements analysis and the feedback of the results of mechanical tests. Chemical analyzes, microstructural analyzes, mechanical tests and finite element analysis were carried out, taking as reference the guidelines established in the ASTM F136-13 and ASTM 2346-11 standards. During the development of this project, several geometric modifications were presented as a result of the constant feedback between the simulation of the design and the behavior of the implant under cyclic loads. Finally, the optimal design of the implant was found and this design achieves the functional requirements of a cervical implant.

Keywords : ASTM F136; ASTM 2346; 3d fatigue behavior; cervical implant; 3d printing; finite element analysis in 3d printing; ELI alloy; Ti6AL4V.

        · abstract in Spanish     · text in Spanish     · Spanish ( pdf )