vol.48 número3Valores de referencia de aminoácidos, acilcarnitinas y succinilacetona por espectrometría de masas en tándem para su uso en el tamizaje neonatal en el suroccidente de ColombiaSíndrome de fragilidad y características socioeconómicas y de salud de los adultos mayores índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Servicios Personalizados

Revista

Articulo

Indicadores

Links relacionados

  • En proceso de indezaciónCitado por Google
  • No hay articulos similaresSimilares en SciELO
  • En proceso de indezaciónSimilares en Google

Compartir


Colombia Médica

versión On-line ISSN 1657-9534

Colomb. Med. vol.48 no.3 Cali jul./set. 2017

https://doi.org/10.25100/cm.v48i3.2173 

Original articles

Noise level in a neonatal intensive care unit in Santa Marta - Colombia.

Angélica Patricia Garrido Galindo1 

Yiniva Camargo Caicedo1 

Andres M Velez-Pereira1 

1 Research Group on Environmental Systems Modeling, Environmental and Sanitary Engineering Program, Universidad del Magdalena, Santa Marta, Colombia.


Abstract

Introduction:

The environment of neonatal intensive care units is influenced by numerous sources of noise emission, which contribute to raise the noise levels, and may cause hearing impairment and other physiological and psychological changes on the newborn, as well as problems with care staff.

Objective:

To evaluate the level and sources of noise in the neonatal intensive care unit.

Methods:

Sampled for 20 consecutive days every 60 seconds in A-weighting curves and fast mode with a Type I sound level meter. Recorded the average, maximum and minimum, and the 10th, 50th and 90th percentiles. The values are integrated into hours and work shift, and studied by analysis of variance. The sources were characterized in thirds of octaves.

Results:

The average level was 64.00 ±3.62 dB(A), with maximum of 76.04 ±5.73 dB(A), minimum of 54.84 ±2.61dB(A), and background noise of 57.95 ±2.83 dB(A). We found four sources with levels between 16.8-63.3 dB(A). Statistical analysis showed significant differences between the hours and work shift, with higher values in the early hours of the day.

Conclusion:

The values presented exceed the standards suggested by several organizations. The sources identified and measured recorded high values in low frequencies.

Key words: Critical Care; Noise Pollution; Benchmark Standard; Neonatal Intensive Care; Environmental Health

Resumen

Introducción:

El ambiente de las unidades de cuidado intensivo neonatal está influenciado por numerosas fuentes de emisión de ruido, que contribuyen a elevar los niveles de ruido y que pueden provocar deficiencias auditivas entre otras alteraciones fisiológicas y psicológicas sobre el neonato, así como problemas al personal asistencial cuando se exceden los niveles.

Objetivo:

Evaluar el nivel y fuentes de emisión del ruido en la unidad de cuidado intensivo neonatal.

Métodos:

Se muestreo durante 20 días continuos cada 60 segundos en escala de ponderación frecuencias A y espacial fast, con un sonómetro tipo I, se registraron los parámetros acústicos de nivel medio, máximo y mínimo, y percentil 10, 50 y 90. Se integraron en periodos horarios y por turno, y se estudiaron mediante un análisis de varianza. Las fuentes se caracterizaron en tercios de octavas.

Resultados:

El nivel medio reportado fue 64.00 ±3.62 dB(A), con máximo de 76.04 ±5.73 dB(A), mínimo de 54.84 ±2.61 dB(A) y ruido de fondo de 57.95 ±2.83 dB(A). Se identificaron cuatro fuentes con rango entre 16.8-63.3 dB(A). El análisis estadístico mostró diferencias significativas entre una media y otra con valores mayores en las primeras horas del día.

Conclusión:

Los valores presentados superan los estándares sugerido por diversas organizaciones. Las fuentes identificadas y medidas registraron mayor aporte en las frecuencias bajas.

Palabras clave: Cuidados críticos; contaminación Sonora; estándares de referencia; cuidado intensivo Neonatal; Salud Ambiental

Introduction

In the environment of the intensive care units, auditory stimuli for the newborns occur daily, frequently associated with the noise caused by the alarms from the medical equipment, telephones, conversations among personnel, closing and opening of doors, and things falling within the unit 1.

These stimuli caused by noise produce four types of adverse effects on newborns, especially among premature newborns, such as somatic effects, sleep disturbances, auditory damage and problems in their emotional development (1,2, as well as the possible repercussions among the care staff 3. Brown stated that excessive auditory stimulation creates negative physiological responses, such as sleep apnea and fluctuations in cardiac frequency, blood pressure and oxygen intake4. It was estimated that noise from voices and monitor alarms can generate an increase in the level of noise around 120 A-weighted decibels, also dB(A), in the units, (5) causing hearing loss, alterations in the newborn’s development, irritability, stress and negative effects on the development of the newborn’s sensory nervous system 6.

Studies have been published that show the average levels of noise in the neonatal intensive care units (NICU). In Chile, values between 45-80 dB(A) were obtained 2; in Lima (Peru), between 62-76 dB(A) 7,8; in Madrid and Huelva (Spain), between 51-88 9,10; in Tabriz (Iran), between 56-70 dB(A) 6; and between 58-70 dB(A) from other studies 8. These are high values when compared with the limits stipulated by the World Health Organization (WHO) that establishes values of 35 dB(A) for the day and 30 dB(A) for the night 11.

In the Colombian Caribbean region, there is a lack of published studies related to the subject of verifying the noise levels that are managed inside a NICU in the region, which is why the equivalent continuous noise levels were assessed in Santa Marta in a hospital’s neonatal intensive care unit.

Materials and Methods

The study was carried out in the neonatal intensive care unit (NICU) of a hospital located in Santa Marta (Colombia), in a public high complexity medical center in the region, which provides medical services while teaching university students in the health area.

The unit has fourteen beds authorized for neonatal critical care, with an average occupation rate of 58% (8±1 per bed) and 5±1 availability of care staff 12 on three shifts: morning (07:00-13:00), afternoon (13:00-19:00) and night (19:00-07:00+1). Furthermore, there are eight feeding schedules per day in 3 hour periods, beginning at 00:00. Finally, in the NICU, there are two, 20 min visiting periods at 11:00 and 17:00.

Sampling design

Noise level in the NICU

The sampling was done continuously for 20 days in the NICU, considering the methodology laid out in the study developed by Vélez-Pereira 13 and Fortes-Garrido et al (10. A Casella type 1 sound level meter, CEL-633-C1K1 model, was used, and it was programmed to record data every 60 seconds using the A frequency weighting filters and the Fast temporary weighting filter. The sound level meter was located in the intensive care unit, taking into account the study developed by Vélez-Pereira 13, the internal dynamic of the ICU and the conditions provided by the care staff coordinators. The sampling point was located in the environment of the unit at a distance of 60 cm from the ceiling and 215 cm from the wall.

The acoustic parameters recorded were the A-weighted equivalent continuous level (LAeq), the maximum A-weighted level (LAmax) and the minimum A-weighted sound level (LAmin), in order to analyze the temporary variation of noise in the Unit. Additionally, the 90th percentile acoustic parameters (LA90) were recorded to establish the background noise of the unit, the 10th percentile (LA10) to establish the dynamic of the peak times or random noises, and the 50th percentile as a contrast of the noise in the unit, which is associated with the existing dynamic among the peak and bottom values of the unit.

Sources of noise emission

A visual identification of the possible sources of emission was done, and then the sound level meter was programmed and a spectral analysis was performed in one-third octave bands without a frequency weighting filter (flat frequencies), which allowed to determine each band’s contribution to the noise levels. The measurement was done within the unit under the influence of other sources of noise emission, justifying the immobility of the equipment mainly due to the dynamic and demand of cubicles in the intensive care unit. The measurements were carried out at a distance of 1.35 m from the sources and at 1.20 m from the level of the floor during three minutes. The same sound level meter, which was used to measure the NICU’s environmental noise, was used.

Processing the Information

Equivalent continuous level of noise in the unit

The Integration of the data recorded of all the noise parameters was done in Microsoft Excel® for two periods, the first corresponding to the care staff shifts (morning, afternoon and night) and the second corresponding to the hourly values. These integration periods were established to determine the hourly variation of the noise throughout the day and the shifts, establishing the possible influence in the dynamics of the noise level in the NICU.

The integrations were done following equation 1.

Equation 1

Where j corresponds to the integrated acoustic parameter, while i corresponds to the number of observations or measurements taken in the integration time interval (hour or shift) and n is the total information observed for the acoustic parameter j in time interval i.

Statistical Analysis

The statistical analysis was executed with the IBM SPSS 20 program. An ANOVA6,13 was performed to compare the averages of the different integration times. Furthermore, the information is confirmed through the Spearman correspondence analysis. Finally, a correspondence analysis was done between the two groups of acoustic parameters (LAeq vs LA50, LAmax vs LA10 and LAmin vs LA90), in order to verify the feasibility of the obtained records.

Results

The hourly LAeq values showed average values between 59.54 ±0.50 dB(A) for 03:00 and 65.27 ±0.46 dB(A) for 08:00, with a maximum value of 69.96 dB(A) (14:00) and a minimum of 57.80 dB(A) (03:00). In Figure 1 (above), the variation of the middle hours of the day can be observed. There, it can be seen that in the early hours of the day (23:00-5:00), the values of the equivalent continuous level of noise were much lower than those presented in the other times.

For the LAmax, values were reported in a range of 67.22 ±3.01 dB(A) (03:00) and 77.32 ±1.52 dB(A) (07:00), with a maximum value of 83.70 dB(A) (23:00) and a minimum value of 59.73 dB(A) (03:00). Finally, for the case of the LAmin, the values varied between 53.16 ±2.45 dB(A) (03:00) and 55.08 ±1.88 dB(A) (08:00), with a maximum and minimum of 61.23 dB(A) and 49.83 dB(A), respectively. The dynamic of these last two parameters was similar to LAeq (Fig. 1, above); notwithstanding, it was seen that the variation of the levels of daily noise was slightly greater in the LAmax values, followed by LAeq and LAmin.

Figure 1 Average level of noise throughout one sampling day for the acoustic parameters. Above: LAeq, LAmax and LAmin; below: percentiles LA50, LA10 and LA90

On the other hand, the noise levels by shift showed that LAeq varied between 64.73 ±1.43 dB(A) and 63.01 ±1.45 dB(A), with a maximum of 68.05 dB(A) and a minimum of 60.27 dB(A). Subsequently for the case of LAmax, values in a range of 74.71 ±1.46 dB(A) to 76.99 ±1.41 dB(A) were present, with maximums and minimums of 80.57 dB(A) and 71.56 dB(A), respectively. Finally, the LAmin varied between 50.54 ±2.1 dB(A) and 54.65 ±1.7 dB(A), with a maximum of 59.58 dB(A) and a minimum of 51.45 dB(A). The maximum values of these ranges in the three parameters were present in the morning shift, the minimums in the night shift, while the absolute maximums and minimums were present in the afternoon, except for the minimum LAmin, which was reported in the night shift (Fig. 2, above). Therefore, it can be affirmed that the acoustic parameters declined in the middle of the night (night shift) and increased in the morning and afternoon shifts. This dynamic was similar to the hourly averages with fewer changes (Fig. 1, above).

Figure 2 Average level of noise by sampling day for the acoustic parameters. Above: LAeq, LAmax and LAmin; below: percentiles LA50, LA10 and LA90

The LA10, LA50 and LA90 acoustic parameters showed that the average times of peak noise (LA10) varied between 62.25±2.39 dB(A) (03:00) and 68.58±1.69 dB(A) (18:00), with a maximum of 73.87 dB(A) and a minimum of 54.76 dB(A). The LA50 varied between 58.39±2.35 dB(A) at 03:00 and 62.85±2.02 dB(A) at 08:00, with a maximum and minimum of 58.00 dB(A) and 51.69 dB(A), respectively. The background noise (LA90) varied less than the two previous parameters (55.51±2.42 dB(A) at 03:00 - 58.74±1.87 dB(A) at 08:00), with a maximum of 63.70 dB(A) and a minimum of 50.61 dB(A) (Fig. 1, below). The hourly fluctuations of the percentiles show that from 22:00 to 04:00 there was a progressive decline, which increased at 05:00 and remained stable for the rest of the day. In terms of the averages by shift (Fig. 2, below), a major variation was observed in LA10 (range 67.94±1.47 - 66.85±1.41 dB(A), maximum 71.31 dB(A), minimum 63.06 dB(A)), followed by LA50 (range 62.30±1.53 - 61.00 ±1.18 dB(A), maximum 65.91 dB(A), minimum 58.64 dB(A)), and ending with LA90 as the most stable parameter (range 58.18 ±1.61 - 57.39 ±1.81 dB(A), maximum 61.88 dB(A), minimum 54.36 dB(A)).

In terms of the emission sources, four sources were identified and characterized, three of which were vital sign monitors of different brands and models and the telephone used for communication (Fig. 3). In general, the results showed similar contributions in the different one-third octave bands. Where the alarms from monitors 1 and 3 had less of a contribution in the low frequencies (from 0 to 125 Hz), while the alarm from monitor 2 and the telephone showed less contribution in high frequencies (greater than 500 Hz).

Figure 3 Spectrum of the studied noise sources. Results from the spectrum of a) A1 sign monitor, b) A2 sign monitor, c) A3 sign monitor, d) telephone 

Promptly, the alarm from monitor 1 varied between 30.6-62.1 dB with a standard deviation of 7.8 dB, presenting a greater value in the high frequencies with an average of 58.09 dB, followed by the middle frequencies (160-400 Hz), with an average of 52.9 dB, and finally with the low frequencies of 51.86 dB (Fig. 3a). This same dynamic was present with the alarm from monitor 3 (Fig. 3c), with a similar variation range (29.3-62.7 dB, standard deviation 7.7 dB), as well as the support by frequencies (high frequencies: 56.65 dB, middle: 56.34 dB, low: 50.46 dB). The alarm from the second monitor (Fig. 3b) contributed more in the low frequencies (50.55dB), followed by the high frequencies (49.20 dB) and middle frequencies (48.36 dB), varying in a range of 17.7-56.6 dB and a standard deviation of 8.8 dB. The telephone spectrum showed an order of frequencies by contribution of low (57.92 dB), middle (43.79 dB) and high (42.91 dB), with a greater variation range (16.8-63.3 dB) and higher standard deviation (12.9 dB). In this last source, three tones marked on three one-third octave bands of a central value of 1,250 and 2,500 Hz were seen (Fig. 3d).

The analysis of variance showed a significant difference between one hourly average and another, with a significance level of 99% for the LAeq and LAmax, while for the LAmin, it did not. In contrast with the ANOVA, the Spearman coefficients confirmed the information, showing a significant relationship with p <0.01 between the hourly averages of LAmax and LAeq, while LAmin showed a slight relationship with p<0.05 (Table 1). These results were similar to the averages by shift, only the Spearman correlation coefficient established an indirect and significant relationship for LAeq and LAmax, and an insignificant relationship for LAmin (Table 1).

Table 1 Analysis of Variance of the noise levels in the NICU of the hospital. 

Acoustic Parameter Statistical Hourly Integration Shift Integration
S F P** SC F p** SC
LAeq 64.00 3.62 10.94 0.00 0.269** 16.44 0.01 -0.381**
LAmax 76.04 5.73 25.04 0.00 0.343** 26.23 0.00 -0.467**
LAmin 54.84 2.61 0.83 0.70† 0.102* 0.03 0.97† -0,032
LA10 67.13 3.95 12.05 0.00 0.273** 6.12 0.00 -0.403**
LA50 61.77 3.16 5.88 0.00 0.205** 2.91 0.06† -0.295*
LA90 57.95 2.83 2.53 0.00 0.134** 0.97 0.39† -0,177

x̄: Average noise level.

S standard deviation.

F statistic from the ANOVA. p statistical meaning for the variance (P-Value from the ANOVA).

SC Spearman correlation coefficient.

* p <0.05.

** p <0.01.

means that there is no significant difference between the average of one level and another with a confidence level of 99%.

For the case of the ANOVA percentiles of the middle hours, a significant difference was seen for the hourly averages middle hours in the three parameters (LA10, LA50 and LA90) with a confidence level of 99% and it is ratified by the Spearman coefficient presented with the same level of meaning (Table 1). In terms of the averages by shift, ANOVA showed only significant differences for LA10, showing that the surrounding and average NICU noise was similar in terms of the shifts. This was confirmed with the Spearman analysis that maintained a significant and indirect relationship with LA10 and LA50 (Table 1).

Finally, the Spearman analysis showed that the noise results are consistent and coherent given that there is a direct and significant correlation between the LAeq-LA50, LAmax-LA10 and LAmin-LA90 variable pairs (Table 2).

Table 2 Correlation of the acoustic parameters in the NICU of the hospital. 

Correlation Hourly Integration Shift Integration
LAeq versus LA50 0.954** 0.973**
LAmax versus LA10 0.890** 0.916**
LAmin versus LA90 0.833** 0.946**

** p<0.01

Discussion

With the obtained results, there were greater records seen during the morning in contrast with those presented in the night, where the values are more constant. The difference can be attributed to medical controls, sampling, radiography, social work and family visits, which occur mostly in the daytime (morning and afternoon). This is confirmed by the hourly integrated averages that showed high values during times around the execution of the mentioned activities and being consequent with the behavior of the percentiles ((LA10, LA50 and LA90), which show a gradual decrease during night shifts. This dynamic has not only been reported for NICUs, but also for any type of ICU 14.

If the relationship between the two groups of acoustic parameters studied is analyzed in the two periods of time, it can be established that the maximum values (LAmax and LA10) and the noise level (LAeq and LA50) show a high variation compared to the minimum (LAmin) and bottom (LA90) values, which do not appear to be influenced by the routine events mentioned, such as feeding and family visits, among others, which is corroborated by the analysis of variance.

Comparing the noise levels reported by the study with the levels recorded by previously published studies, similar values are observed to those presented by different authors in neonatal intensive care units whose values fluctuate between 49-92 dB(A) (6,8,10,13,15-19; a range that includes the average values of LAmin and LAmax and even the average background noise level is greater than the minimum value of the range reported by the research.

The maximum noise levels suggested by the international standards for the protection against acoustic contamination collected by Garrido et al. (20, are exceeded by the results of this study, creating concern about the levels to which newborns and medical personnel are exposed. This situation gets even worse if the suggestion given by the Spanish Pediatrics Association (AEPED, for the Spanish original) is analyzed 21, which suggests that the background noise levels in the unit must not surpass 55 dB(A) and avoid surpassing 70 dB(A), a suggested value exceeded by the results.

These values are more inclusive than those recommended by Konkani and Oakley 11, who maintain that levels less than 40 dB(A) are required for the activities that require concentration, affirming that the greater values can cause interruptions. In the case of the care staff, it is a key limiting factor, given that the nurses must be capable of concentrating on caring for patients in order to avoid causing potential harm due to error 3,11.

On the other hand, when reviewing the results from a spectral footprint and the contribution of each noise band generated by the three monitors and telephone, in two of the four cases it was seen that the bands that contributed on a larger extent to the noise level correspond to the bands located in low frequencies; a result that contributes to increasing the risk of injuring the inner ear of the newborns. Studies show that low frequency sounds are more harmful to the ear’s hair cells, causing microtears, vascular lesions and very frequently hearing loss 22,23.

Various authors report in their studies values between 59-77 dB(A) for the vital sign monitors 6,7,13,24,25, coherent with the results obtained. For their part, the values obtained for the phone in the intensive care unit were less than those recorded by the literature 7,13,24. Notwithstanding, the presence of tones in the high frequencies can induce an abrupt response in the motor skills of the staff and newborns. Furthermore, if the average sound emission sources are compared to the suggestion made by the AEPED, the values exceed the level of 40 dB(A), suggested by this association 21.

The results obtained on the level of noise in the NICU show the need to design methods, strategies and/or programs according to the internal dynamic of the unit, which allow to decrease the noise records, and in this way, lower the risk of creating damaging environmental conditions for the recovery process of the newborns in the NICU.

It is important to note that, among the study’s limitations, the noise emission source measurements in the NICU were not possible to do in totally isolated conditions, due to the demand of the equipment and the dynamics of the unit; additionally, it is declared that it did not include a measuring protocol with a dosimeter in the methodology, the possible risk of exposure will be subject to the time of permanence of staff and/or newborns to the environmental noise levels reported by the sound level meter, which have been documented in other studies and discussed in this study.

Conclusions

In the NICU, a level of noise greater than the limits established by national and international groups was obtained. Also, there was evidence of an influence from the time of day and the shift in the acoustic parameters recorded in the unit. Finally, the sources identified and measured reported greater contributions in low frequencies, associated with a greater risk of harm for the newborn.

References

1. Gallegos-Martínez J, Reyes-Hernández J, Fernández-Hernández EVA, González-González LO. Indice de ruido en la unidad neonatal. Su impacto en recién nacidos. Acta Pediátrica México. 2011; 32(1): 5-14. [ Links ]

2. Fernández DMP. Intervención sensorio-motriz en recién nacidos prematuros. Rev Pediatría Electrónica. 2004; 1(1): 13-20. [ Links ]

3. Garrido GAP, Camargo CY, Vélez-Pereira AM. Nivel continuo equivalente de ruido en la unidad de cuidado intensivo neonatal asociado al síndrome de burnout. Enferm Intensiva. 2015; 26(3): 92-100. [ Links ]

4. Brown G. NICU Noise and the Preterm Infant. Neonatal Netw. 2009; 28(3): 165-73. doi:10.1891/0730-0832.28.3.165. [ Links ]

5. Fernández DP, Cruz JN. Efectos del ruido en ambiente hospitalario neonatal. Cienc Trab. 2006; 8(20): 65-73. [ Links ]

6. Valizadeh S, Hosseini MB, Alavi N, Asadollahi M, Kashefimehr S. Assessment of sound levels in a neonatal intensive care unit in Tabriz, Iran. J Caring Sci. 2013; 2(1): 19-26. [ Links ]

7. Centeno MDV, Apac AA, Sánchez TJC, Raffo NM, Centeno MCA. Niveles de ruido y fuentes asociadas en una Unidad de Cuidado Intensivo Neonatal. Rev Peru Pediatría. 2005;58(1):12-14. [ Links ]

8. Werner F. Evaluación del ruido ambiental en las Unidades de Cuidados Intensivos de Recién Nacidos (UCIN). In: VI Congreso Iberoamericano de Acústica - FIA 2008. Buenos Aires: Universidad Autonoma de Buenos Aires; 2008. [ Links ]

9. Nieto SA. Evaluación de los niveles de ruido en una unidad de cuidados intensivos neonatales. Tesis Doctoral. Facultad de Medicina, Departamento de Pediatría, Universidad Autónoma de Madrid. 2012. Madrid, España. [ Links ]

10. Fortes-Garrido JC, Vélez-Pereira AM, Gázquez M, Hidalgo-Hidalgo M, Bolívar JP. The characterization of noise levels in a neonatal intensive care unit and the implications for noise management. J Environ Health Sci Eng. 2014; 12(1): 104. [ Links ]

11. Konkani A, Oakley B. Noise in hospital intensive care units-a critical review of a critical topic. J Crit Care. 2012; 27(5): 522.e1-9. [ Links ]

12. Vélez-Pereira AM, Camargo CY. Análisis de los factores ambientales y ocupacionales en la concentración de aerobacterias en unidades de cuidado intensivo del Hospital Universitario Fernando Troconis, 2009 Santa Marta - Colombia. Rev Cuid. 2014; 5(1): 595-605.. [ Links ]

13. Vélez-Pereira AM. Evaluación de niveles de ruido en la unidad de cuidados intensivos neonatal del Hospital Juan Ramón Jiménez. Tesis Máster. Máster Oficial en Tecnología Ambiental. Universidad Internacional de Andalucía-Universidad de Huelva. 2010. España [ Links ]

14. Garrido GAP, Camargo CY, Vélez-Pereira AM. Nivel de ruido en unidades de cuidado intensivo de un hospital público universitario en Santa Marta (Colombia). Med Intensiva. 2016; 40(7): 403-10. [ Links ]

15. Berg AL, Chavez CT, Serpanos YC. Monitoring noise levels in a tertiary neonatal intensive care unit. CONTEMPORARY ISS COMMUN SCI DISOR. 2010;37:69-72. [ Links ]

16. Matook S, Sullivan M, Salisbury A, Miller R, Lester B. Variations of NICU sound by location and time of day. Neonatal Netw. 2010; 29(2): 87-95. [ Links ]

17. Alvarez AA, Terrón A, Boschi C, Gómez M. Review of noise in neonatal intensive care units?? regional analysis. J Physics. 2007; 90(1): 012038. [ Links ]

18. Fajardo DL, Gallego SY, Argote LÁ. Noise levels in the Hospital Universitario del Valle newborn intensive care unit Cirena, Cali, Colombia. Colomb Med. 2007; 38(4 Supl 2): 64-71. [ Links ]

19. Yoshiko KT, Moreira PE, Pizzarro G, Guilherme A. Nível de ruído em unidade de terapia intensiva neonatal. Acta Paul Enferm. 2007; 20(4): 404-9. [ Links ]

20. Garrido GA, Camargo CY, Vélez-Pereira AM. Nivel de ruido en la unidad de cuidado intensivo adulto: Medición, estándares internacionales e implicancias sanitarias. Univ Salud. 2015; 17(2): 163-9. [ Links ]

21. García del RM, Sánchez LM, Doménech M E, Izquierdo M I, López H M, Losada M A, et al. Revisión de los estándares y recomendaciones para el diseño de una unidad de neonatología. An Pediatr (Barc). 2007; 67(6): 594-602. [ Links ]

22. Moscoso B. Pérdida auditiva inducida por ruido -PAIR - en Trabajadores del servicio de lavandería del Hospital Arzobispo Loayza. Especialista en Otorrinolaringología. Universidad Nacional Mayor de San Marcos, Lima, Perú. 2003. [ Links ]

23. Maqueda BJ, Ordaz CE, Cortés BRA, Gamo G MF, Bermejo GE, Silva M A, et al. Efectos extra-auditivos del ruido, salud, calidad de vida y rendimiento en el trabajo. Instituto de Salud Carlos III, Escuela Nacional de Medicina del Trabajo: Madrid. 2010. [ Links ]

24. Triló DS, Matos M, Tozo TC, Toso LC, Tomiasi AA, Delfino DPA. Practicing silence: educational intervention for reducing noise in the Intensive Care Unit. Rev Bras Enferm. 2012; 65(2): 285-90. [ Links ]

25. Lawson N, Thompson K, Saunders G, Saiz J, Richardson J, Brown D, et al. Sound intensity and noise evaluation in a critical care unit. Am J Crit Care. 2010; 19(6): e88-e98. [ Links ]

Financing: Project financed by Young Researchers and Innovators Program "Virginia Gutierrez de Pineda" belong to the Administrative Department of Science, Technology and Innovation-Colciencias code 525-2011 , with the support of University of Magdalena.

Financiación: Proyecto financiado a través del Programa Jóvenes Investigadores e Innovadores "Virginia Gutiérrez de Pineda" del Departamento Administrativo de Ciencia, Tecnología e Innovación-Colciencias, código 525-2011 con el apoyo de la Universidad del Magdalena.

Received: December 14, 2015; Revised: March 02, 2017; Accepted: April 05, 2017

Corresponding author: Yiniva Camargo Caicedo. Address: Carrera 32 N° 22 - 08, Santa Marta, Colombia. ZIP Code: 470004. P.O. Box: 2-1-21630. Telephone: +57 (5) 4217940. E-mail: ycamargo@unimagdalena.edu.co

Conflict of interest:

The authors did not present any conflict of interest with the entities that have supported and / or financed this research, as well as with the place where the experimental phase was carried out.

Conflicto de interés:

Los autores no presentaron ningún conflicto de intereses con las entidades que han apoyado y / o financiado esta investigación, así como con el lugar donde se llevó a cabo la fase experimental.

Creative Commons License This is an open-access article distributed under the terms of the Creative Commons Attribution License