vol.41 número1Brote de Salmonella entérica subsp. entérica serovar Give asociado con enfermedad transmitida por alimentos en Vichada, Colombia, 2015Vigilancia por el laboratorio de Shigella spp. aislada de casos clínicos humanos en Colombia, 1997-2018 índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Servicios Personalizados

Revista

Articulo

Indicadores

Links relacionados

  • En proceso de indezaciónCitado por Google
  • No hay articulos similaresSimilares en SciELO
  • En proceso de indezaciónSimilares en Google

Compartir


Biomédica

versión impresa ISSN 0120-4157versión On-line ISSN 2590-7379

Resumen

GARCIA, Karol Julieth et al. Image quality, reading, and mammography service in four diagnostic imaging centers in Manizales, Colombia. Biomed. [online]. 2021, vol.41, n.1, pp.52-64.  Epub 19-Mar-2021. ISSN 0120-4157.  https://doi.org/10.7705/biomedica.5135.

Introduction:

Mammography quality is directly related to the ability to detect an abnormality and, therefore, quality control is necessary for diagnostic imaging centers.

Objective:

To evaluate image quality, reading, and mammography service in some diagnostic imaging centers in Manizales, Colombia.

Materials and methods:

Four diagnostic imaging centers participated voluntarily in the study under confidentiality agreements. Out of 520 women attending the centers, 318 had a mammography. The infrastructure, technology, and human resources of each unit were evaluated based on visual inspections. A radiologist expert in reading and clinical interpretation of mammary images evaluated the quality of the image and the reading. We made the statistical analysis using anova, the kappa index, and the percentage of disagreement.

Results:

We found images of diminished quality mainly due to the presence of artifacts in 75 % of those evaluated, as well as non-compliance with identification criteria and image labeling. There were difficulties in taking the lateral median oblique projection given the absence of the inframammary. The level of agreement in the BI-RADS reporting was low in the four centers with important differences in the report and description of findings.

Conclusion:

The city's diagnostic centers under evaluation are authorized for their operation. However, there are important deficiencies in image quality and reading, which highlights the need to seek quality standards starting from those aspects that can be improved upon.

Palabras clave : Mammography; quality management; diagnostic imaging; quality assurance, health care; maintenance and engineering, hospital; biomedical engineering.