vol.27 número2SUPERVIVENCIA DE LOS PACIENTES CON CÁNCER ESCAMOCELULAR BUCAL, TRATADOS POR PRIMERA VEZ, EN CENTROS ONCOLÓGICOS EN EL PERIODO 2000 A 2011, MEDELLÍN-COLOMBIAEVALUACIÓN DEL GRADO DE MICROFILTRACIÓN EN RESTAURACIONES DE RESINA COMPUESTA, COMPARANDO DOS SISTEMAS ADHESIVOS TRAS DIFERENTES PERÍODOS DE ENVEJECIMIENTO índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Servicios Personalizados

Revista

Articulo

Indicadores

Links relacionados

  • En proceso de indezaciónCitado por Google
  • No hay articulos similaresSimilares en SciELO
  • En proceso de indezaciónSimilares en Google

Compartir


Revista Facultad de Odontología Universidad de Antioquia

versión impresa ISSN 0121-246X

Resumen

CASTRO-ESPITIA, Luis Efrén; NARANJO-RESTREPO, Rodrigo Felipe  y  DUQUE-SERNA, Francisco Levi. MECHANICAL COMPARISON OF TWO INTERNAL RIGID FIXATION SYSTEMS USED FOR FIXING FACIAL FRACTURES (AN IN VITRO STUDY). Rev Fac Odontol Univ Antioq [online]. 2016, vol.27, n.2, pp.262-280. ISSN 0121-246X.  https://doi.org/10.17533/udea.rfo.v27n2a2.

Introduction:

the use of titanium miniplates is the gold standard in the treatment of facial fractures. A Colombian company has developed a system of osteosynthesis miniplates for the treatment of facial fractures, which was chemically and mechanically compared with an imported system.

Methods:

comparative in-vitro tests were carried out to analyze the chemical composition and the biomechanical behavior of the national and imported systems. A sample of 36 items was selected, including 24 miniplates and 12 screws, in order to mechanically assess their materials by means of tensile and flexural tests for the plates and torque strength tests for the screws, and to perform a microscopic analysis of the chemical composition of the plates of both brands by means of scanning electron microscopy (SEM) and Energy Dispersive Spectroscopy (EDS).

Results:

The findings showed that there are no major differences in terms of the chemical composition and the biomechanical behavior of the implants of both brands (national and imported) when assessing plates and screws separately.

Conclusions:

the results showed a higher content of titanium in the material of national plates; however, the mechanical properties of both brands are very similar. Taking into account the material′s biomechanical characteristics, either system is a good choice when selecting a device for rigid internal fixation of facial fractures.

Palabras clave : facial fractures; miniplates; biomechanical evaluation.