vol.48 número3Valores de referencia de aminoácidos, acilcarnitinas y succinilacetona por espectrometría de masas en tándem para su uso en el tamizaje neonatal en el suroccidente de ColombiaSíndrome de fragilidad y características socioeconómicas y de salud de los adultos mayores índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Servicios Personalizados

Revista

Articulo

Indicadores

Links relacionados

  • En proceso de indezaciónCitado por Google
  • No hay articulos similaresSimilares en SciELO
  • En proceso de indezaciónSimilares en Google

Compartir


Colombia Médica

versión On-line ISSN 1657-9534

Resumen

GARRIDO GALINDO, Angélica Patricia; CAMARGO CAICEDO, Yiniva  y  VELEZ-PEREIRA, Andres M. Noise level in a neonatal intensive care unit in Santa Marta - Colombia. Colomb. Med. [online]. 2017, vol.48, n.3, pp.120-125. ISSN 1657-9534.  https://doi.org/10.25100/cm.v48i3.2173.

Introduction:

The environment of neonatal intensive care units is influenced by numerous sources of noise emission, which contribute to raise the noise levels, and may cause hearing impairment and other physiological and psychological changes on the newborn, as well as problems with care staff.

Objective:

To evaluate the level and sources of noise in the neonatal intensive care unit.

Methods:

Sampled for 20 consecutive days every 60 seconds in A-weighting curves and fast mode with a Type I sound level meter. Recorded the average, maximum and minimum, and the 10th, 50th and 90th percentiles. The values are integrated into hours and work shift, and studied by analysis of variance. The sources were characterized in thirds of octaves.

Results:

The average level was 64.00 ±3.62 dB(A), with maximum of 76.04 ±5.73 dB(A), minimum of 54.84 ±2.61dB(A), and background noise of 57.95 ±2.83 dB(A). We found four sources with levels between 16.8-63.3 dB(A). Statistical analysis showed significant differences between the hours and work shift, with higher values in the early hours of the day.

Conclusion:

The values presented exceed the standards suggested by several organizations. The sources identified and measured recorded high values in low frequencies.

Palabras clave : Critical Care; Noise Pollution; Benchmark Standard; Neonatal Intensive Care; Environmental Health.