Introduction
Total intravenous anesthesia (TIVA) is a general anesthesia technique in which only a combination of medications, usually propofol and remifentanyl,1 are administered intravenously.
TIVA is currently administered according to a specific concentration and a desired clinical effect, applying pharmacokinetic models that help predict the desired concentration of the drug. This technique is called open loop or TCI (target-controlled infusion).2,3
Maintaining adequate anesthetic depth is directly related to optimal anesthetic dosage, which translates into administering the amount of anesthetic needed to maintain a surgical procedure. In the case of open-loop TIVA (TCI), its limitation has been demonstrated in certain populations due to the risk of over and/or underdosing.4-6
In response to this scenario, the authors designed a TIVA delivery system where the control of drug perfusion is automatically determined by hemodynamic clinical variables and by the bispectral index (BIS) value.7 The system controller automatically calculates the optimal perfusion velocity based on the current value and the desired value of the control variables, generating stable and fast anesthetic perfusion modifications more accurately. This type of anesthetic control is called a closed loop.7
Although the closed-loop systems in TIVA are not new, very few clinical studies have attempted to compare the performance of both systems, specifically small case series or trials whose outcome is purely technical, but not clinical.8-14
The objective of this study was to evaluate the therapeutic effectiveness of a new closed-loop intravenous anesthesia delivery system to maintain anesthetic depth, compared to an open-loop infusion system controlled by TCI targeting.
Methods
This was a 1:1 randomized, single-blind, 2-parallel groups, controlled superiority clinical trial. The study protocol was submitted for review and approval by the Ethics Committee of the Hospital Universitario San Vicente Fundación (Act 03-2015 of March 13, 2015; registered in clinicaltrials. gov before initiation: NCT02492282).
Participants
Adults over 18 years of age were selected for elective non-cardiac surgery requiring general anesthesia. Pregnant women, surgeries requiring peripheral nerve block before surgery and patients who did not consent to participate were excluded.
Procedure
Before the procedure, the patient was monitored: basic American Society of Anesthesiology (ASA) and electroen-cephalographic monitoring with BIS Vista monitor (Medical Systems, Boston, MA). Two Graseby 3400 perfusion pumps (Graseby Medical, Hertfordshiere, UK) were connected to a venous access. Subsequently, both pumps were connected to the processing and control unit, made up of a personal computer with the application developed by the research team (Fig. 1). This program contained the pharmacokinetic models needed to perform both TCI and closed-loop anesthesia. If the patient was assigned to the intervention group, the controller assumed the anesthetic perfusion according to the BIS, heart rate, and blood pressure, modifying the anesthetic perfusion rate every 5 seconds according to diffuse logic, and using a BIS of 45 as reference values and a heart rate and individualized blood pressure according to the patient. If the patient was assigned to the control group, the anesthesiologist programmed the TCI pumps according to the pharmacokinetic models of Schnider (propofol) and Minto (remifentanyl), and the modifications were made manually according to the patient.
Outcomes
Primary: Therapeutic efficacy determined by the percentage of time spent in an adequate anesthetic depth state (BIS between 40 and 60).15,16
Secondary: Proportion of patients with more than 80% of the time with BIS between 40 and 60; technical performance of the controller; adequate perioperative analgesia with analgoscore -3 to +3. Analgoscore is a scale based on hemodynamic parameters, designed and validated to measure the state of analgesia.12,17 Other outcomes assessed were: manual changes in anesthetic perfusion; change from anesthetic technique to halogenated; hemodynamic instability; intraoperative recall; intraoperative patient movement.
Sample size
A sample size was calculated for a mean difference in the percentage time of the general anesthetic in adequate depth of 10%, a value determined by Hemmerling et al12 in their clinical trial as a reference value to consider a clinically significant mean time difference between 2 intravenous anesthetic delivery systems. With a 2-tailed alpha error of 0.05 and a power of 90%, a sample of 73 per intervention group was obtained. STATA 12.0 was used (Statacorp, College Station, TX).
Statistical analysis
Sociodemographic and baseline clinical characteristics were described by frequencies and percentages for qualitative variables, and with measures of central tendency and dispersion for quantitative variables. For outcomes of a quantitative nature, a mean difference was made with the use of the Student t test, assuming normality in the data provided by the central limit theorem.18 For qualitative variables, relative risks were reported, and Chi-square tests were used to evaluate statistical significance. For each result their respective 95% confidence interval and their 2-tailed P value were reported, with a P of less than 0.05 being considered statistically significant.
The formulas described by Varvel et al19 were used to evaluate the technical performance of the controller:
In this formula, i is the patient number, j is the jth measure of an observation period, and N is the total number of measurements during the observation period. Performance error (PE) is defined as the difference between the actual values and the target value. The median percentage error (MDPE) is a measure of bias and the median absolute percentage error (MDAPE) measures controller inaccuracy. In this context, the Wobble is taken as a measure of intraindividual variability for PE.19
Statistical analyses were performed using STATA 12.0 and SPPS 21.0 (IBM Corporation, Armonk, NY).
Results
The study was conducted between May 2015 and May 2016 (Fig. 2).
No loss during follow-up occurred; however, several interruptions of the assigned intervention were generated in each group (Fig. 2).
The basal clinical and demographic characteristics of the patients are presented in Table 1.
ASA=American Society of Anesthesiology, BP=peripheral nerve block, NSAIDs=nonsteroidal anti-inflammatory drugs, SD=standard deviation, TCI=target-controlled infusion.
Source Authors.
A difference of 15.8%, 95% confidence interval (CI) 9.9 to 21.65, was found between the 2 groups for the mean time spent in an adequate range of anesthetic depth. In addition, relative risk (RR) 2.78 (95% CI: 1.60 4.78) was found in favor of the closed loop for a surgical time greater than 80% in adequate anesthetic depth (Fig. 3). For analgesia and other outcomes, see Tables 2 and 3.
BIS=bispectral index, CI=confidence interval, DP=difference of proportions, MD=mean difference, MDAPE=mean absolute performance error, MDPE=mean performance error, RR=relative risk, TCI=target-controlled infusion.
Source Authors.
* Mean values (SD) are reported.
†An analgesia value of analgoscore between 3 and +3 is considered adequate.
Source: Authors.
CI=confidence interval, MD=mean difference, RR=relative risk, TCI=target-controlled infusion.
* Se report mean values with standard deviation (SD).
Source: Authors.
The change in anesthetic technique within each group was less in closed-loop versus open-loop patients with RR 0.24 (95% CI 0.13 0.47).
For the safety outcomes, it was found that the incidence of hemodynamic instability in TCI was 8% compared to 2.6% in closed loop, whose difference was not statistically significant (P = 0.146). Regarding intraoperative movement, an incidence of 12% was found in the TCI group compared to 10.7% in closed loop (P = 0.79), and none of the groups reported episodes of intraoperative recall.
Discussion
The study demonstrated that the automatic anesthetic delivery system controlled by clinical variables for propofol and remifentanyl was able to maintain up to almost 3 times an adequate state of anesthetic depth based on the measurement of the bispectral index compared to the TCI systems. When evaluating the state of anesthetic depth by categories (Fig. 3), it can be observed that the closed-loop system presents a good to excellent anesthetic control in almost 80% of the patients analyzed, while the TCI system locates almost 60% of the patients in ranges of poor or inadequate anesthetic depth.
Regarding the technical performance of the system, both systems show acceptable performance, with internationally acceptable parameters for these type of devices: 10 to 20% MDPE (bias) and 20 to 40% MDAPE (precision),20,21 and that there is a clear statistically significant difference in favor of the closed-loop system. However, when evaluating the wobble index, which should be the closest to zero,20,21 neither system achieves the objective, which maybe highly related to the inability of both to predict the normal course of a surgical intervention, something that supports the constant need for The presence and performance of the anesthesiologist in a surgical act. These findings are consistent with those reported by other authors.12,14
Another result of the study was the high incidence of changes in the TCI technique (49.3%) versus the need for manual intervention in closed-loop systems (12%), mostly secondary to episodes of intraoperative movement in both groups (12% in TCI vs 10.7% in closed loop). These findings partially question the ability of the BIS to determine the state of anesthetic depth, which may eventually limit the technical and clinical capacity of this type of device (processed indices). In fact, the systematic use of BIS as the only anesthetic depth monitoring method21,22 is increasingly being questioned, and in this study, this situation was evident, given that the main cause of change or modification of the anesthetic technique in both groups (30.7%) was erroneous BIS information about the depth state, which led to erroneous decision-making by the anesthesiologist or the control system.
The biggest limitation was the unblinding of the anesthesiologist who performed the intervention, which could lead to a possible Hawthorne effect. Although this can bear issues of internal validity, having a clinically and statistically significant outcome with more than optimal control strengthens it.
In addition, this study was only able to determine the system's ability to follow certain electroencephalographic parameters, so it is not possible to infer from this an impact on strong clinical results.
The greatest strength of the study lies in the fact that it is the first clinical trial that simultaneously evaluates both the clinical and technical performance of a closed-loop system, using a TCI system as a comparison group.
In conclusion, the closed-loop automatic anesthetic delivery system with the BIS presents a better clinical performance in patients who will undergo intravenous general anesthesia, which does not replace the work of the anesthesiologist, but complements it. However, more research is needed on how best to monitor the anesthetic depth status of the surgical patient.
Ethical responsibilities
Protection of people and animals. The authors declare that for this research experiments have been carried out on human beings, with the prior approval of the Institutional Ethics Committee of the Hospital Universitario San Vicente Fundación, through Act No. 03-2015 of March 13, 2015, and complying with the standards of the Declaration of Helsinki and Resolution 8430 of 1993 regulating clinical and experimental research on patients in Colombia. No animal experiments were performed.
Data confidentiality. The authors state that they have followed their workplace protocols on the publication of patient data.
Right to privacy and informed consent. Authors have obtained informed consent from patients and/or subjects referred to in the article. This document is in the possession of the corresponding author.