Revista Facultad de Odontología Universidad de Antioquia

Print version ISSN 0121-246X

Rev Fac Odontol Univ Antioq vol.33 no.1 Medellín Jan./June 2021  Epub Aug 30, 2021

https://doi.org/10.17533/udea.rfo.v33n1a1 

Original Article

Denture base polymers for analog and digital manufacturing: comparative study of the flexural strength, elastic modulus, and compressive strength of their mechanical properties1 2

Polímeros para fabricación análoga y digital de bases de dentadura: un estudio comparativo de la resistencia flexional, módulo elástico y resistencia a la compresión de sus propiedades mecánicas

Andrés Felipe Vásquez-Niño1  * 

José Rodolfo Ochoa-Alzate2 

Daniel Osorio-Amariles3 

Henry Alberto Rodríguez-Quirós4 

1MSc in Mining, Materials, And Metallurgical Engineering, Federal University of Rio Grande do Sul. Brasil. New Stetic Research Group (GINEWS). https://orcid.org/0000-0002-6456-5583

2Applied Industrial Chemistry Technologist, National Learning Service (SENA). Colombia. New Stetic Research Group (GINEWS). https://orcid.org/0000-0003-0800-4041

3 Pharmaceutical Chemistry Professional, Universidad de Antioquia. Colombia. New Stetic Research Group (GINEWS). https://orcid.org/0000-0002-6227-6307

4 PhD in Chemical Sciences, University of Antioquia. Colombia. New Stetic Research Group (GINEWS). https://orcid.org/0000-0002-4884-6908


ABSTRACT

Introduction:

the emerging manufacture technologies for dental restorations have brought new materials with them, such as 3D-printing resins and CAD/CAM discs for the manufacturing of denture bases. Currently, there is no rigorous mechanical characterization for these materials in the literature, apart from the ones reported in technical datasheets.

Method:

samples for mechanical characterization were manufactured with a conventional heat cure acrylic, a CAD/CAM polymethyl methacrylate (PMMA) disc and two 3D-printing resins. The samples were tested in a universal testing machine, according to ISO 20795-1 for flexural strength and elastic modulus. Compression strength was also determined under dry conditions. The average value of each property was calculated (n = 5). One-way ANOVA and Tukey’s multiple comparisons tests were used.

Results:

mean flexural strengths ranged from 78.35±2.99 to 87.48±4.47MPa, elastic moduli were between 2125.43±57.05 and 2277.72±58.46MPa, and compression strengths values ranged from 85.03±2.14 to 119.15±2.87MPa. Statistical analyses showed significant differences for flexural and compression strengths but did not show any difference for elastic moduli.

Conclusions:

all the tested materials met the minimum required specification for mechanical properties given by ISO 20795-1. From a mechanical point of view, the new materials for digital technologies, i.e., CAD/CAM disc and 3D-printing resins, are suitable for denture-base applications.

Keywords: dental materials; acrylic resins; denture bases; 3d printing; CAD-CAM

RESUMEN

Introducción:

con la aparición de nuevas tecnologías de manufactura han surgido nuevos materiales, como resinas de impresión 3D y discos CAD/CAM, todos empleados para fabricación de bases de dentadura. Actualmente no se cuenta con caracterizaciones mecánicas rigurosas para estos materiales, salvo lo expresado en fichas técnicas.

Método:

se fabricaron muestras para caracterización de propiedades mecánicas con un acrílico termopolimerizable convencional, un disco CAD/CAM de polimetilmetacrilato (PMMA) y dos resinas de impresión 3D. Se fallaron las probetas en una máquina universal de ensayos, según lo exigido por la norma ISO 20795-1 para el caso de la resistencia y módulo de flexión. La resistencia a la compresión también fue determinada. Se calculó el valor promedio de cada propiedad (n = 5). Se realizó un análisis de varianza de una vía y un análisis de Tukey para comparaciones múltiples.

Resultados:

los valores de resistencia a la flexión oscilaron entre 78.35±2.99 y 87.48±4.47MPa; el módulo de flexión estuvo en un rango entre 2125.43±57.05 y 2277.72±58.46MPa. La resistencia a la compresión fluctuó entre 85.03±2.14 y 119.15±2.87MPa. Los análisis estadísticos indicaron diferencias significativas para las resistencias a la flexión y compresión, pero no evidenciaron diferencias para el módulo de flexión.

Conclusiones:

todos los materiales evaluados cumplieron con la especificación mínima de propiedades mecánicas, dada por la ISO 20795-1. Desde el punto de vista mecánico, los nuevos materiales para las tecnologías digitales, discos CAD/CAM y resinas de impresión 3D, son aptos para su aplicación en manufactura de bases de dentadura.

Palabras clave: materiales dentales; resinas acrílicas; bases de dentadura; impresión 3d; CAD-CAM

INTRODUCTION

The emergence and optimization of manufacturing processes using CAD/CAM (computer- aided design/computer-aided manufacturing) and rapid prototyping (3D printing) have enabled its implementation in various technical production areas. Dentistry, in its clinical and laboratory-related components, has not been oblivious to this phenomenon. It is increasingly common to see the use of desktop or intraoral scanners1, specialized design software2, Computerized Numerical Control (CNC) milling machines,3 and 3D laser stereolithography printers or DLP (Digital Light Processing)4 in workflows. The aim of introducing this technology is to replace the stages that characterize analogous workflows, such as taking impressions with elastomeric materials, plaster casting, polymer preparation, and analogous design and manufacturing of restorations or prostheses, among others. Such analogous workflows entail a high dedication of man-hours per job, in addition to the possibility (inherent to the human factor) of making errors in the preparations during the process. Typically, a 100%-digital workflow begins with taking the patient’s impression by means of an intraoral scanner. A digital model is thus obtained in STL (Standard Tessellation Language) format, which is used for designing the restoration using dental CAD software. Finally, the restoration’s design is sent to subtractive (CNC) or additive (stereolithography) manufacturing equipment for manufacturing5,6. These processes guarantee a significant reduction in clinical and laboratory work times, in addition to minimizing errors attributed to the human variable, theoretically obtaining greater reproducibility and precision in the preparations.

Despite the advantages mentioned above, new technologies have introduced other sources of error to the process, including scanners’ precision and resolution, designer skill, number of axes of the CNC milling machine, milling-cutter diameters, equipment calibration, shrinkage of printed parts, positioning of the part for manufacturing, and types of materials for milling and 3D printing. These can make the preparations likely to mismatches or discrepancies1,7,8, impairing the restoration’s performance and, therefore, reducing its lifetime. Out of all the factors mentioned above, the type of material for milling and 3D printing not only influences the restoration’s final fit on the preparation, but it also determines its mechanical properties, which, depending on its final application, can be decisive for adequate clinical performance.

Considering only polymeric materials, the most conventionally used one for provisional- restoration analog manufacturing is PMMA due to its biocompatibility, ease of handling and mechanical performance9. Additionally, this polymer’s mechanical, chemical and tribological properties make it an optimal material to be used in CAD/CAM technology as well, which allow for the manufacturing of crowns, bridges, and denture bases from dense PMMA blocks or discs. Moreover, stereolithography uses acrylic for printing temporary teeth10, denture bases11, clear retainers for invisible orthodontic treatments12, veneers13, etc. Depending on its final application, each resin must have different characteristics and properties that guarantee the ideal performance of the printed piece. In particular, printings made out of denture base resin must meet mechanical, chemical, and physical requirements, regulated by the ISO 20795-1 standard (Dentistry - Base polymers - Part 1: Denture base polymers), being flexural strength and elastic modulus highly important parameters.

Unlike PMMA-based acrylics, the mechanical properties of 3D-printing resins for denture bases have not been thoroughly analyzed. This is because they are commercially new products, and because there is no single formula: each specialized company commercializing this type of material produces its own formula, with a unique composition. Therefore, there is currently no certainty about the mechanical performance of these resins, apart from what the producer specifies in the technical datasheet.

The purpose of this comparative study is to evaluate the differences between the mechanical properties of conventional PMMA, monolayer PMMA blocks used for CAD/CAM processes, and 3D-printing acrylic resins used for denture bases. In order to achieve this, three null hypotheses were proposed:

  • There are no statistically significant differences between the flexural strengths of the evaluated polymers.

  • There are no statistically significant differences between the elastic moduli of the evaluated polymers.

  • There are no statistically significant differences between the compression strengths of the evaluated polymers.

METHODS

Preparation of test specimens

Conventional PMMA

For the manufacturing of the conventional PMMA (PC) test specimens, the Veracril heat- polymerized acrylic (New Stetic S.A.) was used. It was mixed in a 3:1 volumetric proportion with methyl methacrylate (MMA) monomer in a ceramic container. The mixture was left to stand for 10 minutes, which is the time necessary for surpassing the plastic stage according to the product’s use instructions. It was then deposited in mechanical pressing molds for the manufacturing of flexural strength (60x10x3mm) and compressive strength (4mm-diameter and 8mm-height) test specimens. The molds were pressed to 2000 psi during 5 minutes in a laboratory hydraulic press and later they were taken to a thermal reservoir, where a heating curve of 73°C was applied for 90 minutes, and then a curve of 95°C was applied for 30 minutes. Finally, the molds were removed from the reservoir and left outdoors for cooling.

CAD/CAM PMMA disk

A 98.5mm-diameter and 20mm-thick PMMA (CC) monolayer disk of the brand Portux (New Stetic S.A.) was used to manufacture test specimens for flexural (60x10x3mm) and compression strength (4mm diameter and 8mm height) in a Roland DWX-51D milling machine with the aid of the MillBoxLAV software, from STL models (Figure 1A).

Source: by the authors

Figure 1 A. STL designs of flexural and compression strength test specimens B. Manufacturing of flexural strength test specimens by means of stereolithography with the NS resin 

3D resin

Mechanical-characterization test specimens were manufactured in a 405nm ASIGA PICO2 DLP printer from the STL files shown in figure 1A. In order to do that, a 3D-printing resin for denture bases, developed by New Stetic S.A (NS), and a NextDent Base (ND) resin, by NextDent B.V. were used. Both compression and flexural test specimens were printed vertically without supports (Figure 1B), using a layer thickness of 50μm. After the printing process, the parts were removed from the platform and cleaned in ethanol with the aid of an ultrasonic cleaner for 10 minutes. They were then post-cured for 15 minutes in a NextDent LC-3DPrintBox UV curing chamber.

Immediately after preparing the flexural-strength test specimens by applying any of the 3 methods, they were immersed in distilled water at 37°C for 50 hours, as specified by the ISO 20795-1:2013 standard14. Table 1 summarizes the evaluated materials.

Table 1 Evaluated polymeric materials 

Material Description Identification Producer
Veracril Heat-polymerized acrylic PC New Stetic S.A
Portux CAD/CAM disc CC New Stetic S.A
3D denture base 3D resin NS New Stetic S.A
NextDent Base 3D resin ND NextDent B.V

Source: by the authors

Mechanical characterization

Flexural strength and elastic modulus

Flexural-strength tests were carried out according to the ISO 20795-1:2013 standard: After 50 hours of immersion in distilled water at 37°C, test specimens were brought to an INSTRON 4202 universal testing machine, where they were failed inside a water bath at 37°C, using a travel speed of 0.75mm/min. Five test specimens were failed for each manufacturing technique. Flexural strength was determined by using the following expression:

Where F is the maximum force exerted on the test specimen (N), l is the distance between the specimen supports (mm), b is the specimen width (mm), and h is the specimen thickness (mm).

The following equation was used to calculate the elastic modulus:

Where F1 is the force exerted on the test specimen to a deflection of 2mm (N), l, b and h correspond to the distance of the supports, and the width and thickness of the test specimen, respectively (mm), and d is the deflection in millimeters to force F1, i.e., 2mm.

Compressive strength

Compression-strength test specimens were immersed in distilled water for 24 hours at 37°C immediately after being manufactured. The compression strength test was performed in dry conditions, using a universal INSTRON 4202 testing machine, and a travel speed of 0.5mm/ min, at a temperature of 23°C. Compression strength was determined according to the following expression:

Where F is the maximum force applied to the test specimen (N), and d is the diameter of the test specimen (mm).

Statistical analysis

The collected data were analyzed descriptively with a one-way analysis of variance (ANOVA). Additionally, Tukey’s multiple test was used to determine significant differences between groups (α = 0.05). The analyses were performed by using the statistical package STATGRAPHICS Centurion XVI version 16.2.04.

RESULTS

Table 2 presents the average value of the evaluated resins’ flexural strength, elastic modulus, and compression strength, with their respective standard deviation. The average flexural strength values obtained were 78.35, 81.99, 83.51 and 87.48MPa for PC, CC, NS and ND polymers, respectively. The ANOVA analysis (Table 3) showed significant differences between the groups’ flexural strength values (P> = 0.01). Tukey’s test for multiple comparisons (Table 4) confirmed statistically significant differences between the conventional polymer (PC) and NextDent Base resin (P < 0.05). However, no differences were evident among the CC, NS and ND groups.

Table 2 Average values and standard deviation (SD) determined for flexural strength, elastic modulus, and compression strength of the evaluated materials. Additionally, the values reported in the respective technical data sheets (D.S.) are included 

Material Flexural strength (MPa) Elastic modulus (MPa) Compression strength (MPa)
Average SD D. S. Average SD D. S. Average SD D. S.
PC 78.35 2.99 73.29 2224.41 168.08 2277 86.29 0.99 N.R.
CC 81.99 4.01 >80 2125.43 57.05 > 2000 85.03 2.14 N.R.
NS 83.51 3.17 N.R. 2208.76 66.89 N.R. 119.15 2.87 N.R.
ND 87.48 4.47 84 2277.72 58.46 2383 109.42 1.65 N.R.

N.R.: No information reported

Source: by the authors

Table 3 One-way ANOVA analysis for the specific mechanical properties 

Test Origin Sum of squares Degrees of freedom Mean square F P
Between the groups 214.44 3 71.48
Flexural strength Within the groups 220.12 16 13.76 5.20 0.0107
Total 434.56 19
Between the groups 59716.10 3 19905.4
Elastic modulus Within the groups 157595 16 9849.7 2.02 0.1515
Total 217311 19
Between the groups 4336.65 3 1445.55
Compression strength Within the groups 65.92 16 4.12 350.87 0.0000
Total 4402.58 19

Source: by the authors

Table 4 Comparison in pairs using the Tukey’s test. Asterisks (*) indicate statistically significant differences (Sig. diff.) between the evaluated pairs 

Groups Flexural strength Elastic modulus Compressive strength
Sig. diff. Difference +/- Limits Sig. diff. Difference +/- Limits Sig. diff. Difference +/- Limits
PC-CC -3.646 6.71464 98.976 179.664 1.260 3.67448
PC-NS -5.160 6.71464 15.646 179.664 * -32.856 3.67448
PC-ND * -9.134 6.71464 -53.310 179.664 * -23.128 3.67448
CC-NS -1.514 6.71464 -83.330 179.664 * -34.116 3.67448
CC-ND -5.488 6.71464 -152.286 179.664 * -24.388 3.67448
NS-ND -3.974 6.71464 -68.956 179.664 * 9.728 3.67448

Source: by the authors

Similarly, the average values of the elastic moduli obtained for the studied groups were 2224.41 (PC), 2125.43 (CC), 2208.76 (NS) and 2277.72MPa (ND). Neither the analysis of variance nor Tukey’s yielded significant differences between the evaluated groups.

Compression strength was higher than 80MPa in all cases, and 3D-printing resins (NS and ND) had values above 100MPa. Tukey’s test determined that there are significant differences between all evaluated groups, except between the conventional polymer and the CAD/CAM one.

DISCUSSION

In this work, the mechanical properties, and more precisely, the flexural strength, elastic modulus, and compression strength of 4 polymeric materials for denture-base manufacturing were determined. Based on analyses of variance, which demonstrated significant differences between the groups evaluated in terms of flexural and compression strength, null hypotheses were rejected for these mechanical properties, while the null hypothesis for the elastic modulus was accepted.

Dental prostheses must meet biocompatible, aesthetic, and mechanical requirements. The properties evaluated in this study are considered to be of great importance for the proper performance of dental prostheses, since they are subjected to flexural and compressive loads during bite actions and masticatory processes. Additionally, bending overload fracture is clinically considered to be the most common failure of denture bases15. Therefore, it should be ensured that prostheses do not have plastic deformations or fractures during their use in the mouth. Thus, high values of stiffness, flexural and compression strength16 are desired. Flexural strength, determined in a universal testing machine under wet conditions and at body temperature allows for the simulation of the material’s response to the mechanical loads to which it will be subjected during the masticatory process. At the same time, the elastic modulus indicates the stiffness or resistance offered by the material to being irreversibly deformed. Both properties are regulated by the ISO 20795-1 standard, which requires minimum values of 65MPa in flexural strength and 2000MPa in elastic modulus for heat-polymerized polymeric materials that are expected to be used for the manufacturing of denture bases. Consequently, from a mechanical point of view, all materials characterized in this study can be used as materials for such application. Nonetheless, ISO 20795-1 does not specify those values for photopolymerized polymers, such as 3D-printing resins, and thus far, there is no standard for regulating such materials. For this reason, the mechanical requests for heat-polymerized acrylics are assumed to assess the suitability of 3D resins for the above-mentioned application. On the other hand, compression strength is not a normative property according to ISO 20795-1, at least for this application, but it must be considered for this purpose, given that in areas such as the posterior teeth, the prosthesis is subjected to compressive efforts17.

The polymers analyzed in this work can be classified into two large groups according to their chemical nature: The first one includes PC and CC materials, both being heat-polymerized PMMA polymers. Specifically, CAD/CAM blocks or disks are manufactured with PMMA using compression, and are polymerized using a heating curve, i.e., similarly to the conventional works on acrylic. The other group consists of 3D-printing resins, NS and ND, which, as mentioned above, are photopolymerized liquid acrylics. On the one hand, PMMA resins are monofunctional, linear-chain compounds, resulting in lower flexural strength and elastic modulus18. This was not evidenced in the mechanical characterization. Apart from the PC and ND comparison, the difference between the properties of the evaluated materials was not significant. This is due to the fact that PC and CC materials are reinforced with a crosslinking agent in order to obtain branched and/or crosslinked chains, which restricts their movement and results in an increase in mechanical resistances19. On the other hand, 3D-printing resins for dental applications are composed of monomers and acrylic oligomers that can be mono-, bi-, and even trifunctional, such as bisphenol A-glycidil dimethacrylate or urethane dimethacrylate20. This favors obtaining a highly crosslinked and, therefore, stiffer, and more resistant polymer. Similarly, it can be seen in Table 2 that the experimental values correspond to those specified in the corresponding product’s technical datasheets21-23, which apparently validates the information provided by businesses. However, no technical information reported by manufacturers was found for the compression strength of the evaluated polymers. This is because this property is not regulated by ISO 20795-1. Similarly, there is no technical information available for New Stetic 3D resin, as this is a product currently undergoing the development phase, and it is not yet being sold by the company.

The mechanical properties of 3D-printed polymers do not depend solely on their composition and chemical structure. Variables like printing parameters and post-curing process affect the mechanical performance of the pieces manufactured through this technology20. Additionally, stereolithography is an additive manufacturing technique, which means that the printing of pieces is made through the continuous overlapping of layers with certain thickness. The test specimens for flexure and compression made with NS and ND resins were printed vertically (Figures 1B and 2A), which means that the direction of the uniaxial load in the flexion tests was parallel to the layers of the test specimens, whereas the direction of the compression load was perpendicular (Figure 2B). Alharbi et al.25 assessed the influence of the printing direction of cylindrical test specimens in the compression strength of a resin used for temporary teeth. It was observed that vertically printed test specimens (layers perpendicular to the direction of the load) had significantly higher compression strength than horizontally printed test specimens (layers parallel to the direction of the load). This occurs because the load applied parallel to the layers that compose the test specimen induces the separation or delamination of these more easily than when oriented in a perpendicular direction. This would suggest an increase in the flexural strength of test specimens made with NS and ND resins if printed in such a way that the layers are perpendicular to the direction of the uniaxial load during the test. However, this would mean increasing the area of contact between layers, which could increase the parts’ degree of porosity, residual stress, and dimensional distortions by shrinkage26, thus counteracting the possible increase in mechanical resistance.

Source: by the authors

Figure 2 A and B. Arrangement of printing layers and layer orientation with respect to the uniaxial direction of the load of mechanical flexure and compression tests 

Although the mechanical results obtained in this study are promising for 3D-printing resins, it is necessary to carry out subsequent preclinical and clinical studies where the mechanical resistance of dental prostheses, printed in different orientations and positions, is evaluated for the application of more complex flexure, tensile and compressive loads. This would allow the results to be extrapolated in a more assured manner to the actual application of such products.

CONCLUSIONS

The elastic modulus, flexural and compression strength of 4 commercial polymeric materials were determined for the manufacture of denture bases by analog or digital techniques. The findings of this work have led us to conclude:

  • All evaluated materials had strengths and elastic moduli greater than the minimum required by ISO 207095-1. Therefore, if evaluated only from a mechanical point of view, the materials are appropriate to be used for the manufacturing of denture bases

  • 3D-printing resins had higher compression strength values than PMMA acrylics, while there were no statistically significant differences between the evaluated materials either in the strength or in the elastic modulus

  • It is necessary to evaluate the influence of the orientation of the layers of the 3D-printed test specimens on the mechanical properties of the materials.

REFERENCES

1. Rudolph H, Salmen H, Moldan M, Kuhn K, Sichwardt V, Wöstmann B et al. Accuracy of intraoral and extraoral digital data acquisition for dental restorations. J Appl Oral Sci. 2016; 24(1): 85-94. DOI: https://dx.doi.org/10.1590%2F1678-775720150266Links ]

2. Son K, Lee WS, Lee KB. Prediction of the learning curves of 2 dental CAD software program. J Prosthet Dent. 2019; 121(1): 95-100. DOI: https://doi.org/10.1016/j.prosdent.2018.01.004Links ]

3. Boitelle P, Mawussi B, Tapie L, Fromentin O. A systematic review of CAD/CAM fit restoration evaluations. J Oral Rehabil. 2014; 41(11): 853-74. DOI: https://doi.org/10.1111/joor.12205Links ]

4. Galante R, Figuereido-Pino CG, Serro AP. Additive manufacturing of ceramics for dental applications: a review. Dent Mater. 2019; 35(6): 825-46. DOI: https://doi.org/10.1016/j.dental.2019.02.026Links ]

5. Sadid-Zadeh R, Katsavochristou A, Squires T, Simon M. Accuracy of marginal fit and axial wall contour for lithium disilicate crowns fabricated using three digital workflows. J Prosthet Dent. 2020; 123(1): 121-27. DOI: https://doi.org/10.1016/j.prosdent.2018.11.003Links ]

6. Kim DY, Jeon JH, Kim JH, Kim HY, Kim WC. Reproducibility of different arrangement of resin copings by dental microstereolithography: evaluating the marginal discrepancy of resin copings. J Prosthet Dent. 2017; 117(2): 260-65. DOI: https://doi.org/10.1016/j.jds.2017.06.006Links ]

7. Kang SY, Lee HN, Kim JH, Kim WC. Evaluation of marginal discrepancy of pressable ceramic veneer fabricated using CAD/CAM system: additive and subtractive manufacturing. J Adv Prosthodont. 2018; 10(5): 347-53. DOI: https://dx.doi.org/10.4047%2Fjap.2018.10.5.347Links ]

8. Schaefer O, Kuepper H, Sigusch BW, Thompson GA, Hefti AF, Guentsch A. Three-dimensional fit of lithium disilicate partial crowns in vitro. J Dent. 2013; 41(3): 271-7. DOI: https://doi.org/10.1016/j.jdent.2012.11.014Links ]

9. Bombac D, Brojan M, Fajfar P, Kosel F, Turk R. Review of materials in medical applications. Materials and Geoenvironment. 2007; 54(4): 471-99. [ Links ]

10. Digholkar S, Madhav VNV, Palaskar J. Evaluation of the flexural strength and microhardness of provisional crown and bridge materials fabricated by different methods. J Indian Prosthodont Soc. 2016; 16(4): 328- 34. DOI: https://doi.org/10.4103/0972-4052.191288Links ]

11. Kalberer N, Mehl A, Schimmel M, Müller F, Srinivasan M. CAD-CAM milled versus rapidly prototyped (3D-printed) complete dentures: an in vitro evaluation of trueness. J Prosthet Dent. 2019; 121(4): 637-43. DOI: https://doi.org/10.1016/j.prosdent.2018.09.001Links ]

12. Cole D, Bencharit S, Carrico CK, Arias A, Tüfekçi E. Evaluation of fit for 3D-printed retainers compared with thermoform retainers. Am J Orthod Dentofacial Orthop. 2019; 155(4): 592-9. DOI: https://doi.org/10.1016/j.ajodo.2018.09.011Links ]

13. Al-Rimawi A, EzEldeen M, Schneider D, Politis C, Jacobs R. 3D printed temporary veneer restoring autotransplanted teeth in children: design and concept validation ex vivo. Int J Environ Res Public Health. 2019; 16(13): 496-505. DOI: https://dx.doi.org/10.3390%2Fijerph16030496Links ]

14. International Organization for Standardization. ISO 20795-1: 2013. Dentistry - Base polymers - Part 1: denture base polymers. 2nd ed. 35p, 2013. [ Links ]

15. Chitchumnong P, Brooks SC, Stafford GD. Comparison of three-and four-point flexural strength testing of denture base polymers. Dent Mater. 1989; 51(1): 2-5. DOI: https://doi.org/10.1016/0109-5641(89)90082-1Links ]

16. Ajaj-ALKordy NM, Alsaadi MH. Elastic modulus and flexural strength comparisons of high-impact and traditional denture base acrylic resins. Saudi Dent J. 2014; 26(1): 15-8. DOI: https://doi.org/10.1016/j.sdentj.2013.12.005Links ]

17. Sepideh B, Saman S. In vitro comparative study of compressive strength of different types of composite resins in different periods of time. Iran J Pharm Sci. 2008; 4(1): 69-74. [ Links ]

18. Haselton DR, Diaz-Arnold AM, Vargas MA. Flexural strength of provisional crown and fixed partial denture resins. J Prosthet Dent. 2002; 87(2): 225-8. DOI: https://doi.org/10.1067/mpr.2002.121406Links ]

19. Saen-Isara T, Dechkunakorn S, Anuwongnukroh N, Srikhirin T, Tanodekaew S, Wichai W. Influence of the cross-linking agent on mechanical properties of PMMA powder with compromised particle morphology. Int Orthod. 2017; 15(2): 151-64. DOI: https://doi.org/10.1016/j.ortho.2017.03.007Links ]

20. Revilla-León M, Meyers MJ, Zandinejad A, Özcan M. A review on chemical composition, mechanical properties, and manufacturing work flow of additively manufactured current polymers for interim dental restorations. J Esthet Restor Dent. 2019; 31(1): 51-7. DOI: https://doi.org/10.1111/jerd.12438Links ]

21. New Stetic. Ficha técnica: resina acrílica termopolimerizable Veracril®, Opt i-cryl®. DPFTPT-025. Guarne, Antioquia: New Stetic; 2020. Documento de referencia: DPDDPR-019. Available in https://bit.ly/3g30HaFLinks ]

22. New Stetic. Ficha técnica: discos PMMA para CAD/CAM PORTUX HCN. DPFTPT-087. Guarne, Antioquia: New Stetic ; 2020. Documento de referencia: DPDDPR-019. Available in https://bit.ly/2Oyh0RdLinks ]

23. 3D Systems Next Dent [Internet]. Netherlands: 3D Systems; 2019. Available in https://nextdent.com/products/denture-3dplus/Links ]

24. Puebla K, Arcaute K, Quintana R, Wicker RB. Effects of environmental conditions, aging, and build orientations on the mechanical properties of ASTM type I specimens manufactured via stereolithography. Rapid Prototyping Journal. 2012; 18(5): 374-88. [ Links ]

25. Alharbi N, Osman R, Wismeijer D. Effects of build direction on the mechanical properties of 3D-printed complete coverage interim dental restorations. J Prosthet Dent. 2016; 115(6): 760-7. DOI: https://doi.org/10.1016/j.prosdent.2015.12.002Links ]

26. Dimitrov D, Schreve N, de Beer N. Advances in three dimensional printing - state of the art and future perspectives. Rapid Prototyping Journal . 2006; 12(3): 136-47. DOI: https://doi.org/10.1108/13552540610670717Links ]

1 This work was partially supported by the National Learning Service (SENA) in the context of the call for submissions for the Promotion of Innovation and Technological Development in Companies, with a regional approach.

2Vásquez-Niño AF, Ochoa-Alzate JR, Osorio-Amariles D, Rodríguez Quirós HA. Denture base polymers for analog and digital manufacturing: comparative study of the flexural strength, elastic modulus, and compressive strength of their mechanical properties. Rev Fac Odontol Univ Antioq. 2021; 33(1): 6-16. DOI: http://dx.doi.org/10.17533/udea.rfo.v33n1a1

Received: June 11, 2019; Accepted: November 03, 2020

*CORRESPONDING AUTHOR Andrés Felipe Vásquez, afvasquez@newstetic.com, Carrera 53 #50-09, (+57) 550 0000 Ext. 1504, New Stetic S.A Guarne, Colombia.

CONFLICTS OF INTEREST

The authors declare that there is no conflicts of interest.

Creative Commons License This is an open-access article distributed under the terms of the Creative Commons Attribution License